Description
The Fluke 393 FC CAT III 1500 V True-rms Solar Clamp Meter with iFlex is designed for solar photovoltaic (PV) installation technicians and maintenance professionals who work in high voltage DC environments: PV arrays, wind power, electric railways, data centers battery banks for uninterruptible power supplies. The clamp meter will measure up to 1500 V DC and 1000 V AC with test leads, and up to 999.9 A DC or AC through the clamp jaw. The included iFlex flexible current probe extends AC current measurements up to 2500 A. The 393 has a thin jaw, giving you access to cables in crowded combiner boxes. Plus, the test leads are designed with your work in mind, and are also rated to CAT III 1500 V DC.
Features
- IP54 rated, ideal for work outdoors including PV panel testing
- DC power measurement, showing readings in kVA
- Audio Polarity indicator helps prevent accidental miswires
- Visual Continuity provides a bright green light in the display, ideal when working in dark and noisy environments
- Logging and reporting of test results via Fluke Connect software
- When measuring ac current the included iFlex flexible current probe gives you unparalleled access to cable is tight spaces. The iFlex probe can be twisted through extremely small spaces and provide accurate current measurements.
Specifications
Specifications: | |||
General | |||
Maximum voltage between any Terminal and Earth Ground | |||
AC | 1000 V | ||
DC | 1500 V | ||
Batteries | 2 AA IEC LR6 alkaline | ||
Display | Dual display with backlight | ||
Automatic Power Off | 20 minutes | ||
Electrical | |||
Accuracy | |||
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, relative humidity at 0 % to 75 %. Accuracy specifications take the form of: ±([% of Reading] + [Number of Least Significant Digits]). | |||
Temperature Coefficients | Add 0.1 x specified accuracy for each °C > 28 °C or < 18 °C | ||
AC Current: Jaw | |||
Range | 999.9 A | ||
Resolution | 0.1 A | ||
Accuracy | 2 % + 5 digits (10 Hz to 100 Hz) | ||
2.5 % + 5 digits (100 Hz to 500 Hz) | |||
Crest Factor (50/60 Hz) | 2.5 @600.0 A | ||
3.0 @500.0 A | |||
1.42 @999.9 A | |||
Add 2 % for C.F. >2 | |||
AC Current: Flexible Current Probe | |||
Range | 999.9 A | ||
2500 A | |||
Resolution | 0.1 A (≤999.9 A) | ||
1 A (≤2500 A) | |||
Accuracy | 3 % RD + 5 digits (10 Hz to 500 Hz) | ||
Crest Factor (50/60Hz) | 2.5 @1400 A | ||
3.0 @1100 A | |||
1.42@2500 A | |||
Add 2 % for C.F. >2 | |||
Position Sensitivity | |||
Distance from Optimum | i2500-10 Flex | i2500-18 Flex | Error |
A | 0.5 in (12.7 mm) | 1.4 in (35.6 mm) | ±0.5 % |
B | 0.8 in (20.3 mm) | 2.0 in (50.8 mm) | ±1.0 % |
C | 1.4 in (35.6 mm) | 2.5 in (63.5 mm) | ±2.0 % |
Measurement uncertainty assumes centralized primary conductor at optimum position, no external electrical or magnetic field, and within operating temperature range. | |||
DC Current | |||
Range | 999.9 A | ||
Resolution | 0.1 A | ||
Accuracy | 2 % RD + 5 digits[1] | ||
[1] When using the ZERO (B) function to compensate for offsets. | |||
AC Voltage | |||
Range | 600.0 V | ||
1000 V | |||
Resolution | 0.1 V (≤600.0 V) | ||
1 V (≤1000 V) | |||
Accuracy | 1 % RD + 5 digits (20 Hz to 500 Hz) | ||
DC Voltage | |||
Range | 600.0 V | ||
1500 V | |||
Resolution | 0.1 V (≤600.0 V) | ||
1 V (≤1500 V) | |||
Accuracy | 1 % RD + 5 digits | ||
mV dc | |||
Range | 500.0 mV | ||
Resolution | 0.1 mV | ||
Accuracy | 1 % RD + 5 digits | ||
Amps Frequency: Jaw | |||
Range | 5.0 Hz to 500.0 Hz | ||
Resolution | 0.1 Hz | ||
Accuracy | 0.5 % RD + 5 digits | ||
Trigger Level | 5 Hz to 10 Hz, ≥10 A | ||
10 Hz to 100 Hz, ≥5 A | |||
100 Hz to 500 Hz, ≥10 A | |||
Amps Frequency: Flexible Current Probe | |||
Range | 5.0 Hz to 500.0 Hz | ||
Resolution | 0.1 Hz | ||
Accuracy | 0.5 % RD + 5 digits | ||
Trigger Level | 5 Hz to 20 Hz, ≥25 A | ||
20 Hz to 100 Hz, ≥20 A | |||
100 Hz to 500 Hz, ≥25 A | |||
Voltage Frequency | |||
Range | 5.0 Hz to 500.0 Hz | ||
Resolution | 0.1 Hz | ||
Accuracy | 0.5 % RD + 5 digits | ||
Trigger Level | 5 Hz to 20 Hz, ≥5 V | ||
20 Hz to 100 Hz, ≥5 V | |||
100 Hz to 500 Hz, ≥10 V | |||
DC Power | |||
Range | 600.0 kVA (600.0 V dc range) | ||
1500 kVA (1500 V dc range) | |||
Resolution | 0.1 kVA | ||
1 kVA | |||
Accuracy | 2 % RD + 2.0 kVA | ||
2 % RD + 20 kVA | |||
Resistance | |||
Range | 600.0 Ω | ||
6000 Ω | |||
60.00 kΩ | |||
Resolution | 0.1 Ω (≤600.0 Ω) | ||
1 Ω (≤6000 Ω) | |||
0.01 kΩ (≤60.00 kΩ) | |||
Accuracy | 1 % RD + 5 digits | ||
Capacitance | |||
Range | 100.0 μF | ||
1000 μF | |||
Resolution | 0.1 μF (≤100.0 μF) | ||
1 μF (≤1000 μF) | |||
Accuracy | 1 % RD + 5 digits | ||
Inrush Trigger Level | 5 A | ||
Mechanical | |||
Size (L x W x H) | 281 mm x 84 mm x 49 mm | ||
Weight (with batteries) | 520 g | ||
Jaw Opening | 34 mm | ||
Flexible Current Probe Diameter | 7.5 mm | ||
Flexible Current Probe Cable Length | |||
(head to electronics connector) | 1.8 m | ||
Environmental | |||
Operating Temperature | -10 °C to 50 °C | ||
Storage Temperature | -40 °C to 60 °C | ||
Operating Humidity | Non condensing (<10°C) | ||
≤90 % RH (at 10 °C to 30 °C) | |||
≤75 % RH (at 30 °C to 40 °C) | |||
≤45 % RH (at 40 °C to 50 °C) | |||
Operating Altitude | 2000 m | ||
Storage Altitude | 12 000 m | ||
Ingress Protection (IP) Rating | IEC 60529: IP54 non-operating | ||
Electromagnetic Compatibility (EMC) | |||
International | IEC 61326-1: Portable, Electromagnetic Environment, IEC 61326-2-2 CISPR 11: Group 1, Class A | ||
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the internal function of the equipment itself. | |||
Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-voltage power supply network that supplies buildings used for domestic purposes. There may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances. | |||
Caution: This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments. | |||
Korea (KCC) | Class A equipment (Industrial Broadcast & Communications Equipment) | ||
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and not to be used in homes. | |||
USA (FCC) | 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103. | ||
Safety | |||
General | IEC 61010-1, Pollution Degree 2 | ||
Measurement | IEC 61010-2-032: CAT III 1500 V / CAT IV 600 V | ||
IEC 61010-2-033: CAT III 1500 V / CAT IV 600 V | |||
Wireless Radio | |||
Radio frequency certification | FCC ID: T68-FBLE, IC: 6627A-FBLE | ||
Wireless Radio Frequency Range | 2400 MHz to 2483.5 MHz | ||
Output Power | <100 mW | ||
SIMPLIFIED EU DECLARATION OF CONFORMITY | |||
Hereby, Fluke declares that the radio equipment contained in this Product is in compliance with Directive 2014/53/EU. | |||
The full text of the EU declaration is available at the following Internet address: | |||
www.fluke.com/en-us/declaration-of-conformity |
Downloads
Refund Policy
Returns & Exchanges Policy
RETURNING THE PURCHASE
- Refunds on Cash On Delivery (COD) orders will be credited to your MMT eWallet, which can be used for future purchases. Future purchases will be subject to the same COD/shipping fee procedures, if applicable.
- Credit Card returns will be credited back to your original credit card.
- Value Added Tax (VAT) on the product purchased will be refunded.
- If an item you ordered is unavailable, we will notify you by email as soon as possible. If payment has already been made, refunds will be credited back to the original payment card used for the transaction.
- MMT eWallet credits have an expiry period of 12 months.
ORGANISING RETURNS OR EXCHANGES FOR REGISTERED CUSTOMERS
- To organize a return or exchange, registered users must log in to their registered account. Navigate to ‘My Account’ and then to ‘Returns & Exchanges’ to submit your request. The request for a refund must be made within 14 days of receiving your order. You will receive a notification upon completing the request.
- You can track your return or exchange by clicking on ‘Track Your Order’.
ORGANISING RETURNS OR EXCHANGES FOR GUEST USERS
- Guest users must submit a request via email to info@mmt.com within 14 days of receiving their order. Alternatively, you can contact Customer Care at +966-548907080 from Saturday to Thursday , 8:00 am to 8:00 pm (GMT+4). During Ramadan, our working hours are from 8:00 am to 8:00 pm, Monday to Friday.
ADDITIONAL INFORMATION
- Returned items must be received within 14 days of receiving the full order. Returns received after this period will not be processed and will be returned at your expense.
- If you placed an order with multiple items, you can only request a return once all items from your order have been delivered.
- Return or exchange of items is only permissible once.
- Any additional items returned that were not included in the approved list will be sent back to you, and shipping costs will be charged.
- All items must be returned in one package and in their original packaging with all tags attached. This includes any belts or designer packaging included with the item. If items are not returned in their original state, or are used and without tags, they will be sent back to you, and you will incur shipping charges. You will not be allowed to choose your shipping method; returns will be processed through our designated shipping companies only.
- All items are quality checked before dispatch. If you receive a damaged or faulty item, please log in to your registered account, go to ‘My Account’, and follow the steps within 48 hours of receiving your order. Guest users can submit a request via the ‘Track Your Order’ page.
- You will receive an email confirming that your return or exchange item(s) has been received, and the refund will be processed accordingly.
MMT RIGHTS
- MMT reserves the right to refuse refunds or exchanges on items deemed unfit for resale.
- Returns made outside the 14-day timeframe will not be accepted and will be returned to you at your expense.